Gentier48  Real-Time PCR System

Your ideal choice for qPCR

Thermal Block

Sample capacity(wells): 48

Reaction Volume: 5-100ul

Consumables: 0.2ml 8-strip tubes, 0.2 ml PCR single tube, optical flat cap clear tube.

Temperature range: 0°C~100°C

Heating/cooling method: Peltier

Max heating rate: 8,0°C/sec Average heating rate: 6,0°C/sec

Max cooling rate: 5.5°C/sec Average cooling rate: 4,0°C/sec

Temperature accuracy: ±0.1°C

Temperature uniformity; ±0.1°C

Gradient Range: 1°C to 40°C (Modell E + R)

Gradient block:  12 row (Modell E + R)

Heat Lid

Temperature range: Room temperature-110°C

Optical System

Detector: Photodiode

Detection Position: Excitation and scan at top

Detection method: 6channels scanning at the same time, no edge effect

Detection time: 7 seconds for 96 wells for all channels

Range of excitation/emission wavelengths(nm):

1.             465/510(FAM,SYBR Green I, SYTO9, EvaGreen, LCGreen)

2.             527/563(HEX,VIC,TET,JOE)

------------------------------------------------------------

3.             580/616(ROX,Texas Red)

4.             632/664(Cy5)

Probe:Taqman Probe, Molecular beacons probe, Scorpion probe

Multiplexing: Up to 2/4 tragets

Fluorescence Linearity: r≥0.990

Fluorescence Dynamic Range: Adjustable

Performance

Sample Linearity: /r/≥0.999

Sample repeatability: Ct value CV≤0.5%

Software Functions

Data Analysis modes: Qualitative analysis, Absolute quantification, Relative quantification, Genotyping analysis, Endpoint analysis, melt curve analysis, High Resolution Melting.

Control modes:

1. 7' touch screen

2. PC direct control

3. WLAN control(One PC can control max 10 units, and device can be controled bu any PC in the WLAN)

Data Storage: Upload and download through USB disk, 1000 results can be stored in machine

Power failure protection: Automaticly start running experiments after power supply, no need wait PC software.

Customize Report: Templetes reserved, report can customized.

Administration Management: Administrator can set functions limits for users.

Transport Locker: Automaticly detect transport locker

Fault Management: Fault report and analysis, solution instruction

LIS connectuib: CSV, Excel, TXT format data output, open port for LIS connection

Others

PC operating system: Win7, Win 10

Communication Port: 1 ethernet and 2 USB

Footprint (WxDxH): 260mmX400mmX260mm

Weight: 11KG

Power usage: AC100 to 125V/200 to 240 V(50/60 Hz)

Power consumption: 600VA

Work environment: Temperature: 10°C~30°C   Humidity: 20%~85%.



Gentier 96 Real-Time PCR System

Your ideal choice for qPCR

Thermal Block

Sample capacity(wells): 96

Reaction Volume: 0-100ul

Consumables: 0.2ml 96-well Plates(unskirt); 0.2ml 8-strip tubes, 0.2 ml PCR single tube.

Temperature range: 0°C~100°C

Heating/cooling method: Peltier

Max heating rate: 6.1°C/sec Average heating rate: 4.5°C/sec

Max cooling rate: 5.0°C/sec Average cooling rate: 2.8°C/sec

Temperature accuracy: ±0.1°C

Temperature uniformity; ±0.1°C

Gradient Range: 1°C to 40°C (Modell E + R)

Gradient block:  12 row (Modell E + R)

Heat Lid

Temperature range: Room temperature-110°C

Optical System

Detector: Photodiode

Detection Position: Excitation and scan at top

Detection method: 6channels scanning at the same time, no edge effect

Detection time: 7 seconds for 96 wells for all channels

Range of excitation/emission wavelengths(nm):

1.             465/510(FAM,SYBR Green I, SYTO9, EvaGreen, LCGreen)

2.             527/563(HEX,VIC,TET,JOE)

3.             580/616(ROX,Texas Red)

4.             632/664(Cy5)
------------------------------------------------------------

5.             680/730(Alexa Fluor680)

6.             465/616(FRET)

Probe:Taqman Probe, Molecular beacons probe, Scorpion probe, FRET

Multiplexing: Up to 4/6 tragets

Fluorescence Linearity: r≥0.990

Fluorescence Dynamic Range: Adjustable

Performance

Sample Linearity: /r/≥0.999

Sample repeatability: Ct value CV≤0.5%

Software Functions

Data Analysis modes: Qualitative analysis, Absolute quantification, Relative quantification, Genotyping analysis, Endpoint analysis, melt curve analysis, High Resolution Melting.

Control modes:

1. 10.4' touch screen

2. PC direct control

3. WLAN control(One PC can control max 10 units, and device can be controled bu any PC in the WLAN)

Sample Drawer: Touch screen commend

Data Storage: Upload and download through USB disk, 1000 results can be stored in machine

Power failure protection: Automaticly start running experiments after power supply, no need wait PC software.

Customize Report: Templetes reserved, report can customized.

Administration Management: Administrator can set functions limits for users.

Transport Locker: Automaticly detect transport locker

Fault Management: Fault report and analysis, solution instruction

LIS connectuib: CSV, Excel, TXT format data output, open port for LIS connection

Others

PC operating system: Win7, Win 10

Communication Port: 1 ethernet and 3 USB

Footprint (WxDxH): 355mmX480mmX485mm

Weight: 30KG

Power usage: AC100 to 125V/200 to 240 V(50/60 Hz)

Power consumption: 900VA

Work environment: Temperature: 10°C~30°C   Humidity: 20%~85%.



TL988-II Real-Time PCR System

Key Feature

1. Block: 48 samples

2. Easy-to-use Software

3. Various Operation modes

4. Advantage Temperature Control

5. Non-Maintenance Optical Systems

6. Sensitive and Precise Thermoelectric Systems

7. Particular Momentary Power Failure Protection

 

Specifications

Thermal Block

Block formats: 48-well

Sample volume: 10-100 μL

Consumables: Thick-walled 0.2ml PCR tibes, 0.1ml 8-strip tubes

Max heating/cooling rate: 3.0°C/sec/ 2.0°C/sec

Temperature range: 4.0-99.0°C

Temperature accuracy: ±0.1ºC

Temperature uniformity: ±0.3°C

Temperature steps: 1-99

Heated lid

Temperature range: 100-120°C

Optics

Excitation: 2 LEDs

Excitation range: 470-523 nm

Detector: PMT

Detection range: 525-564nm

Detection channels:

Channel     Excitation   Emission     Pre-Calibrated Dyes

1.                   470 nm        525 nm         FAM, SYBR Green I

2.                   523 nm        564 nm         VIC, HEX, TET, JOE

Detection range: 101-1010 copies

Software

Analysis modes: Absolute quantification, Relative quantification,

                                Melt curve analysis

Performance

Reproducibility: CV≤2.1%

Correlation coefficient of nucleic acid precision: R≥0.997

System

Operating systems: Windows 7

Communication: USB

Power usage: AC220 × (1±10%)V, 50 × (1±2%)H

Dimensions: 520W×420D×320H mm

Weight: 27kg



TL988-IV Real-Time PCR System

Your ideal choice for qPCR

TL988 is a effective PCR device for qualitative or quantitative detection of nucleic acids, genotyping, and mutation analysis. With a high-performance yet low cost configurations to accommodate basic to advanced experimental applications. Featuring four optical channels with user-selected filters covering a broad range of excitation wavelengths, it will meet the needs of a broad range of scientific applications in genomics research and as a better partner of your qPCR.

 

Design for flexible use

The TL-988 system adapts to almost any workflow requirement with flexible instrument control and data management.

  • Standard with 7’’ LCD touch screen, functions quickly accessible through simple, one-touch commands
  • High-throughput up to 96 samples per run
  • Push-pull, electrical controlled elevating platform
  • Concealed lid with constant temperature and isopiestic pressure bring high accurate temperature control
  • Quick-start set-up so you can begin a run immediately
  • Unique power-off protection function can save all the setted configurations for sudden power outages, and allow the experiment continues when power supply is restored.
  • System can be installed in three distinct configurations, providing unmatched flexibility and convenience.
    • Co-located configuration (PC control)
    • Standalone (PC-free) configuration (Data can be export by USB and analysis on PC)
    • Networked configuration (PC software automatically detect all systems in LAN and allow to remote monitoring one of experimental progress and downloading of the completed run file to the PC at your desk)

Fast and uniform heating and heat sinking block

  • Concealed lid with constant temperature and isopiestic pressure
  • Electrical controlled elevating & automatic locked platform
  • Thermo block is based on improved heat pipe technology, leading to extraordinary Iintra-well temperature homogeneity and maximized inter-well, inter-cycle reproducibility.
  • Highly efficient PID&DSP control algorithm ensures faster respond and stronger interference resistance, fast and uniform heating and heat sinking

High sensitive and precise optical system

  • High-sensitivity and wide dynamic range of fluorescence detection system, sample dynamic range could be from 101 to 1010 copies
  • Fast and precise mechanical scanning structure ensures the illumination uniformity of 96 wells, Eliminate edge effects, no need for ROX calibration
  • Fluorescence excitation by a high-intensity and long-life LED, Bulb Luminance will not attenuation over time, free maintenance optical system through the entire life.
  • Professional designed fluorescence filters realize the highly fluorescence transmittance and meanwhile minimize the fluorescence channel crosstalk.

Specifications

 

Thermal Block

Block formats: 96-well

Sample volume: 5–50 μL

Consumables: 0.1ml 96-well unskirted Plates; 0.1ml 8-strip tubes (Optical flat cap)

Max heating/cooling rate: 4.2°C/sec

Temperature range: 4.0–99.0°C

Temperature accuracy: ≤±0.1

Temperature uniformity: ≤ ±0.3°C

Temperature steps: 1-99

Heated lid

Temperature range: 100–120°C

Control: Automatic temperature and pressure setting

Optics

Excitation: 4 LEDs

Excitation range: 470–630 nm

Detector: Photodiode

Detection range: 525–670 nm

Detection channels:

Channel    Excitation     Emission      Pre-Calibrated Dyes

1.                 470 nm           525 nm          FAM, SYBR Green I

2.                 523 nm           564 nm          VIC, HEX, TET, JOE

3.                 571 nm           621 nm           ROX, TEXAS-RED

4.                 630 nm          670 nm           CY5

Multiplex: Up to 4 targets

Detection range: 101 -1010 copies

Software

Analysis modes: Absolute quantification, Relative quantification,

                                Melt curve analysis, High Resolution Melting

Performance

Reproducibility: CV≤2.1

Correlation coefficient of nucleic acid precision: R≥0.997

System

Operating systems: Windows 7

Communication: Ethernet or USB

Power usage: AC220 × (1±10%)V, 50 × (1±2%)Hz

Dimensions: 540W×300D×480H mm

Weight: 35kg

Gewährleistung: 1 Jahr